    Johanna Spangler
    Participant

    Hallo Harrison

    Is it possible to measure the OD with Scattering via LED without stopping the Stirrer?
    I don’t know if the Stirrer stops for the OD Measurement via Scattering, its for me not apparent from the Supplementary Data1.
    Is there any way to measure while stirring? And how can i rewrite the program ?
    Do you have any ideas?

    Best regards

    #1581
    harrison
    Keymaster

    Hello Johanna
    Yeah, basically you could just adjust the main loop in the code to start the stirring BEFORE doing OD measurement. Or, add a new line to do a second OD measurement after the stirring has started. It wont cause any issues with the running of the device, however it will make the measurements more noisy particularly if you are doing some violent stirring since there will be so many bubbles + vortex in the way!