I may use the ChiBio at some point, but my cells grow in a very expensive medium.
I was wondering if there is a minimal volume for the instrument to work.
I imagine that lowering the volume could affect both measurement precision and dilution accuracy.
Any feedback on this would be very appreciated.
Thank you for your time,
Emanuele
Hello,
The minimal volume is ~15ml. If you take the sides off the device you can see why this is – the liquid needs to be high enough to be seen by the sensor/laser/LED.